Patient Treatment Guide: Defibrillators | FDA Issues Nationwide Recall of Medtronic Sprint Fidelis Defibrillators

Medtronic Defibrillator Lead Recalls

Medtronic Letter to Patients

Important Patient Information
Sprint Fidelis® Lead Performance

October 15, 2007

Dear Medtronic Heart Device Patient:

Medtronic recently provided doctors with important information about our Sprint Fidelis defibrillation leads (Models 6930, 6931, 6948, 6949). We have informed doctors who, according to our records, manage the care of patients with Sprint Fidelis leads. A defibrillator system consists of a device implanted near the shoulder and one or more leads connecting the device to the heart. This information applies to Sprint Fidelis leads only, not implanted defibrillation or pacemaker devices. If you have a Sprint Fidelis lead, your Patient ID card should contain one of the following 4 sets of numbers: 6930, 6931, 6948, 6949. These numbers may be shown at the beginning of a longer set of numbers on your ID card.

We have found that there is a small chance of fractures in particular locations on the Sprint Fidelis lead. If you have a Sprint Fidelis lead, the chance there is a problem with your lead is small. Patients with a Sprint Fidelis lead are more likely to experience complications from removal than from a problem with a Sprint Fidelis lead. However, doctors may choose to change the way a device is programmed to help reduce any potential problems. An independent panel of physician experts recommends against removing Sprint Fidelis leads except in very unusual circumstances.

If you have a Sprint Fidelis lead, you may have already heard from your doctor’s office regarding this matter. We encourage patients with Sprint Fidelis leads to contact their doctor’s office with any questions about their leads or the information in this letter. The information we provided doctors is more detailed, since they are in the best position to know how to manage patient care in light of specific health considerations.

We sincerely apologize for any difficulties this may cause you and your family. We are committed to answering your questions, keeping you informed, and continuously improving our products. If you have additional questions or would like to inquire if you have a Sprint Fidelis lead, please contact Medtronic Patient Services at 1 (800) 551-5544 x41835 (7am to 9pm CST).

Sincerely,
Reggie Groves
Reggie Groves
Vice President, Quality and Regulatory
Medtronic Cardiac Rhythm Disease Management
Medtronic, Inc.


Medtronic Dear Doctor Letter

Urgent Medical Device Information Sprint Fidelis
Lead Patient Management Recommendations

October 15, 2007

Dear Doctor,

This letter provides important information on Sprint Fidelis lead performance and recommendations for ongoing patient management. Our records indicate that you have implanted or are following patients with Sprint Fidelis leads (Models 6930, 6931, 6948, 6949). In consultation with our Independent Physician Quality Panel, we are voluntarily suspending distribution of Sprint Fidelis leads worldwide. This decision is based on a variety of factors detailed in this letter that when viewed together, indicate that suspension of implantation is the appropriate action. You should no longer implant Sprint Fidelis leads, and you should return any unused product to Medtronic.

Background
As we reported in March 2007, there are two primary locations1 where chronic conductor fractures have occurred on Sprint Fidelis leads: 1) the distal portion of the lead, affecting the anode (ring electrode) and 2) near the anchoring sleeve tie-down, predominantly affecting the cathode (helix tip electrode), and occasionally the high voltage conductor. High voltage conductor fractures could result in the inability to deliver defibrillation therapy. Anode or cathode conductor fractures (at either location) may present clinically as increased impedance, oversensing, increased interval counts, multiple inappropriate shocks, and/or loss of pacing output. The potential for defibrillation lead fracture to result in or contribute to inappropriate therapies or death has been previously reported.2 As of October 4, 2007, there have been approximately 268,000 Sprint Fidelis leads implanted worldwide. Based on current information, we have identified five patient deaths in which a Sprint Fidelis lead fracture may have been a possible or likely contributing factor. We have confirmed 665 chronic fractures in returned leads. Approximately 90% of these fractures have occurred in the anode or cathode conductors, while 10% have occurred in the high voltage conductors.

Performance Update
Since our March 21st communication, we have examined six months additional Returned Product Analysis (RPA) and Medtronic System Longevity Study (SLS) data. In addition, we have performed extensive analysis using the Medtronic CareLink® Network (25,000 devices). These data give us confidence in our current understanding of Sprint Fidelis’ performance.

RPA of Sprint Fidelis leads shows a survival of 99.2% at 30 months. However, RPA overstates actual performance since it does not account for leads that are not returned. The Medtronic SLS data for the Model 6949 Sprint Fidelis lead indicate 97.7% [+1.3/-3.0] all-cause lead survival at 30 months. This is consistent with our analysis of Medtronic CareLink Network data from approximately 25,000 Sprint Fidelis leads, which indicate 97.7% [+0.6/-0.8] survival at 30 months. These survival rates are not statistically different from the all-cause lead survival of 99.1% [+0.4/-0.8] for the Model 6947 Sprint Quattro® lead at 30 months from the SLS. However, we expect this difference will become statistically significant over time if the current failure rates remain constant.

Recommendations
Medtronic recommends you consider the following as part of routine follow-up for each patient

The patient management recommendations set forth above should increase the likelihood that a fracture will be detected by Patient Alert and/or Medtronic CareAlert notifications and decrease the likelihood of inappropriate therapies. Based on our review of the available data, there does not appear to be a benefit to more frequent follow-up.

Medtronic’s Independent Physician Quality Panel believes it is inappropriate to prophylactically remove Sprint Fidelis leads except in unusual individual patient circumstances.We support this position.

Lead extraction carries risks that should be considered in patient management. Published literature suggestsmajor complications (death or surgical intervention) from lead extraction range from 1.4-7.3%.3,4 As always, with confirmed lead failure the risk of extraction should be weighed against the risk of adding an additional lead.

Additional Communication
The HRS-recommended Physician Device Advisory Notice for this communication is attached. The information in this letter will be posted on Medtronic.com on October 15th. Consistent with the HRS5 recommendations on device advisory communications we will be informing patients with affected devices, advising them to contact you for more information. The patient letter will be sent on October 22nd.

We are notifying regulatory agencies of this communication. We will continue to provide performance updates every six months via our Product Performance Report.

Nothing is more important to Medtronic than patient safety. We are committed to answering your questions and keeping you informed. We regret any difficulties this may cause you and your patients. If you have questions or concerns, please contact your Medtronic Representative or Medtronic Technical Services at 1(800) 723-4636 (US).

Sincerely,

Reggie Groves
Vice President, Quality and Regulatory
Medtronic Cardiac Rhythm Disease Management

1 The two primary locations described above account for 90% of the chronic fractures identified by RPA. The remaining 10% of chronic fractures occurred in DF-1 connector leg and the proximal portion of the RV coil.

2 Kleemann T, Becker T, Doenges K, et al., K. Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years. Circulation. May 15, 2007; 115(19): 2461 - 2463.

3 Byrd CL, Wilkoff BL, et al. Intravascular extraction of problematic or infected permanent pacemaker leads: 1994-1996. U.S. Extraction Database, MED Institute. PACE May 2000; 23(5): 927-928.

4 Bracke FA, Meijer A, vanGelder LM. Lead extraction for device related infections: a single centre experience. Europace, May 2004; 6(3): 243-247.

5 Recommendations from the HRS task force on device performance policies and guidelines. Carlson MD, et al. Heart Rhythm Journal 2006, 3, 1250-73.